The spectrophotometer is used for technological control of raw materials and finished products in the process of autoclave leaching of rare-earth metals.
Technical specifications:
- The spectral range is from 190 to 1000 nm;
- Spectral gap width 4 nm;
- Accuracy of setting the wavelength is not more than ± 1 nm;
- Wavelength reproducibility: ± 0,5 nm;
- The limits of the permissible absolute error in the measurement of the spectral transmission coefficients of directional transmission ± 1.0% T (190 - 315 nm) and ± 0.5% T (315 - 1000 nm);
- Measurement range:
Optical density: 3.000 to 0.000
Coefficient of directional transmission: from 0.0 to 100.0 % - Modes of operation: basic / quantitative;
- Working length of the cuvette: 5 - 100 mm;
- Light source: deuterium and halogen lamps;
- Digital output for connection to a PC for managing, analyzing and displaying information via an installed program.