Technical specifications:
- Input of potentiometric sensors;
Measuring range ± 2000 mV
Resolution 0.1 mV
The margin of error of 0.2 mV - Polarized sensors input;
Measuring range 0 ... 2000 mV / 0 ... 200 μA
Resolution 0.1 mV / 0.1 μA
The margin of error of 0.1 mV / pA 0.3 - Temperature sensor PT1000;
Resolution 0.1 °С
The limit of 0.2 ° C error - Dosing module (burette drive);
Burette resolution 1 / 20000 burette volume
For the burette 10 ml - 0.5 μl
The margin of error of 0.3% of the metered volume